​Model T320

with NumaView™ Software

Gas Filter Correlation N2O Analyzer​​​​

​​

Features

 

  • Ranges: 0-1 ppm to 0-1,000 ppm, user selectable
  • Dual ranges and auto ranging
  • Large, vivid, and durable color graphics display with touch screen interface
  • Ethernet, RS-232, and (optional) USB com ports
  • Front panel USB connections for peripheral devices and firmware upgrades
  • 8 analog inputs (optional)
  • Adaptive signal filtering optimizes response time
  • Temperature & pressure compensation
  • Comprehensive internal data logging with programmable averaging periods
  • Ability to log virtually any operating parameter
  • Two-year warranty
  • Five-year warranty on GFC wheel

Maintenance

​The following table provides a list of items that  need regular maintenance to ensure your product remains in good working condition.

Item

Action

Frequency

Cal Check Required

Particulate Filter

Replace

Weekly or As Needed

No

Verify Test Functions

Record and Analyze

Weekly or after any Maintenance or Repair

No

Pump Diaphragm

Replace

Annually

Yes

Perform Flow Check

Check Flow

Annually

No

Perform Leak Check

Verify Leak Tight

Annually or after any Maintenance or Repair

No

Pneumatic lines

Examine and Clean

As Needed

Yes if cleaned

Cleaning

Clean

As Needed

Only if cover removed​


Downloads

  
  
  
SAL000056B - T320.pdfT320 Specification1112 KB
T SERIES_6PG.BROCHURE_2018_web.pdfT Series with NumaView Brochure9567 KB
T320-T320U ADDENDUM_07406.pdfT320-T320U Addendum288 KB

Specifications


Ranges Min: 0 - ​1 ppm full scale
Max: 0 - 1,000 ppm full scale
(selectable, dual-range supported)​
Measurement Units ppb, ppm, μg/m3, mg/m3 (selectable)
Zero Noise < 0.02 ppm​ (RMS)
Span Noise < 0.5% of reading (RMS) above 5 ppm
Lower Detectable Limit 0.04 ppm​
Zero Drift < 0.1 ppm/24 hours
Span Drift < 0.5% of reading/24 hours
Lag Time <10 seconds
Rise/Fall Time < ​​60​ seconds to 95%
Linearity 1% of full scale
Precision 0.5% of readin​​​g above 50 ppb
Sample Flow Rate 800 cc/min ±10%
Power Requirements 100V-120V, 220V-240V, 50/60 Hz
Analog Output Ranges 10V, 5V, 1V, 0.1V (selectable)
Recorder Offset ±10%
Included I/O 1 x Ethernet: 10/100Base-T
2 x RS232 (300-115,200 baud)
2 x USB device ports
8 x opto-isolated digital outputs
6 x opto-isolated digital inputs
4 x analog outputs
Optional I/O 1 x USB com port
1 x RS485
8 x analog inputs (0-10V, 12-bit)
4 x digital alarm outputs
Multidrop RS232
3 x 4 - 20mA current outputs
Operating Temperature Range 5 - 40ºC 
Dimensions (HxWxD) 7” x 17” x 23.5” (178 x 432 x 597 m​m)
Weight


Specifications are subject to change without notice.
All specifications are based on constant conditions.​

50 lbs (16.2 kg)

​​

Description

The Model T320 N2O analyzer measures Nitrous Oxide by comparing infrared energy absorbed by a sample to that absorbed by a reference gas according to the Beer-Lambert law.  Using a Gas Filter Correlation Wheel, a high-energy IR light source is alternately passed through a N2O filled chamber and a chamber with no N2O present.  The light path then travels through the sample cell, which has a folded path of 2.56 meters.

The energy loss through the sample cell is compared with the reference signal provided by the filter wheel to produce a signal proportional to concentration, with little effect from interfering gases within the sample.  This design produces excellent zero and span stability and high signal to noise ratio, allowing excellent performance over a wide concentration range.

All T Series instruments offer an advanced color display, capacitive touch screen, intuitive user interface, flexible I/O, and built-in data acquisition capability.  All instrument set up, control and access to stored data and diagnostic information is available through the front panel, or via RS232, Ethernet, or USB com ports, either locally or by remote connection.

The Model T320 comes with the Numaview™ Software. NumaView™ Remote PC Software allows for a remote connection with virtual interface and data downloading capability to analyzers operating NumaView™ Software.​