​Model T700H

with NumaView™ Software

Source Monitoring Dilution Calibrator​

Features


  • 满足或超过方案205的要求
  • 杰出的气体浓度现场测量灵活性
  • 用于转化器效率检查的气相滴定,能够产出高达100 ppm的NO2
  • 能够减少气体和气缸使用,本金返还快
  • 第三个质量流量控制器以提供更宽广的自动量程和气体混合​(可选)​
    • 经久耐用、宽大生动的彩色触屏展示​
    • 以太网, RS-232, 和(可选的) USB通讯端口
    • 前面板USB连接,可用于外部设备连接和固件升级
    • 12 个独立的序列计时器
    • 嵌套序列(最多5级)
    • 质量流量控制器的软件线性
    • 4个校准端口(可为一种或多种混合气体配置)​​​
  • Two-year warranty

Maintenance

​The following table provides a list of items that  need regular maintenance to ensure your product remains in good working condition.

Item

Action

Frequency

Manual Section

Test Functions

Verify, record and analyze

Weekly or after any Maintenance or Repair

3.3.4
5.1
Appendix C

Dryer Desiccant

Replace

Every 30 hours of O3 Generator run-time*

8.2.1

Flow Check

Verify Flow of MFCs

Annually or any time the T700H’s internal DAC is recalibrated

7.1 & 7.2

Leak Check

Verify Leak Tight

Annually or after any Maintenance or Repair

8.2.2

Pneumatic lines

Examine and clean or replace

As needed (requires cal check after cleaning or replacement)

 - - -

*See Table 9‑1 in manual for details on the desiccant maintenance warning.



Downloads

  
  
  
07464A MANUAL, T700H OPERATOR'S.pdfT700H Manual6330 KB
SAL000072B_T700H.pdfT700H Specification134 KB
T SERIES_6PG.BROCHURE_2018_web.pdfT Series with NumaView Brochure9567 KB

Specifications


 
​Dilution System
Flow Measurement Accuracy ± 1% of full scale

Repeatability of Flow Control ± 0.2% of full scale
Linearity of Flow Measurements ± 0.5% of full scale
Flow Range of Dilution Air 0 - 10 SLPM
Optional Ranges 0 to 20 SLPM, 0 to 5 SLPM
Flow Range of Cylinder Gases 0 to 1 SLPM
Optional Ranges 0 to 5 SLPM
Zero Air Requirements 10 SLPM @ 30 psi
Optional Zero Air Range
20 SLPM @ 30 psi
​​     Input Pressure
​20-40 psig
Calibration Cylinder Gas Input Port 4 (configurable)
Diluent Gas Input Port

1
​NO2 Generation (GPT modes)
Concentration Range 5-100 ppm
Maximum Output (NO2 Generation) GPT Modes 400 ppm LPM
Minimum Output (NO2 Generation) GPT Modes 10 ppm LPM
​​Ozone Generator Module
Maximum Ozone Concentration

100 ppm
Minimum Ozone Concentration 5 ppm
Response Time 180 seconds to 98%
​​​​Electrical Specifications
Power Requirements
85V - 264V, 47Hz - 63Hz
Analog Output Ranges 10V, 5V, 1V, 0.1V (selectable)
​​Communication Specifications
Included I/O

1 x Ethernet: 10/100Base-T
2 x RS232 (300-115,200 baud)
2 x USB device ports
12 x digital control outputs
12 x digital control inputs
8 x digital status outputs
Optional I/O 1 x USB com port
1 x RS485
Multidrop RS232
​​Physical Specifications​
Operating Temperature Range

5 - 40ºC
Dimensions (HxWxD) 7” x 17” x 24” (178 x 432 x 609 mm)
Weight  39.29 lbs (17.83 kg) ​​​​​


Specifications are subject to change without notice.
All specifications are based on constant conditions.



Applications

Description


T700H是基于微处理器的、为连续排放监测系统的校准仪提供校准的稀释校准仪。它能够大大减少气缸的使用和花费,并能够提高场地上气体浓度测量和转化器效率检查的灵活性。该仪器的设计能够满足美国EPA的要求。

T700H能够精准的稀释高浓度气缸中的协议1或通过认证的气体,以提供任何所需浓度的气体。该仪器同样配有稳定、可调节的臭氧发生器和气相滴定以生产二氧化氮。气相滴定功能能够使用一氧化氮校准气源自动生成二氧化氮。产生的二氧化氮浓度可高达100ppm,能够用于连续排放监测系统的NOx分析仪的跨度检查或转化器效率测试。通过配置可选的第三个质量流量控制器,T700H可以混合两个不同气缸中的气体。

所有的T系列仪器提供先进的彩色显示、电容式触屏、直观的用户界面、灵活的I/O和内置的数据采集能力。所有的仪器安装、控制、数据和诊断信息访问都可通过NumaView™在前面板、RS232、以太网或USB通讯端口获得。这些连接可以本地完成,也可通过NumaView™远程电脑软件实现。

T700H安装有NumaViewtm™软件。NumaView™远程电脑软件能够远程连接运行NumaView™的仪器的虚拟接口并下载数据。