​Model T500U

 ​CAPS NO2 Analyzer

T320U 

Features

  • Direct NO2 measurement
  • Extended maintenance interval
  • No converter or reagent required
  • No known ambient air interferences
  • Fast response
  • Exceptional measurement sensitivity
  • Low power consumption
  • Ranges: 0-5 ppb to 0-1 ppm, user selectable
  • Adaptive signal filtering optimizes response time
  • Temperature & Pressure compensation
  • US EPA Federal Equivalent Method (EQNA-0514-212)
  • EN14211, EN15267
  • MCerts: Sira MC160304/00
  • All standard T-Series platform features
  • Available with premier NumaView™software upon request
  • Two-year warranty

Maintenance

​The following table provides a list of items that  need regular maintenance to ensure your product remains in good working condition.

Item Action Frequency Cal Check Manual Section
1Particulate filter Change particle filter Weekly No 10.3.1
Verify test functions Review and evaluate Weekly No 10.2; Appendix C 
Zero/span check Evaluate offset and slope Weekly --  9.3, 9.6, 9.9
1Zero/span calibration Zero and span calibration Every 3 months --  9.2, 9.4, 9.5, 9.7, 9.8
1External zero air scrubber (optional) Exchange chemical Every 3 months No 10.3.3
1Perform flow check Check Flow Every 6 Months No 10.3.7
Internal IZS Permeation Tube  Replace Annually Yes 10.3.2
Perform pneumatic leak check Verify Leak Tight Annually or after repairs involving pneumatics Yes 10.3.6
2Pump diaphragm Replace Annually Yes  Refer to diaphragm kit instructions
 Calibrate UV Lamp Output Perform lAMP CAL  Prior to zero/span calibration or PMT hardware calibration --  5.9.6 & 11.7.2.5
3PMT sensor hardware calibration Low-level hardware calibration On PMT/ preamp changes  if Yes 11.7.2.8
0.7 < slope or
slope >1.3
1Sample chamber optics  Clean chamber, windows and filters As necessary Yes 11.7.2.2 & 11.7.2.3
1Critical flow orifice & sintered filters Replace As necessary Yes 10.3.4

 

1 These Items are required to maintain full warranty; all other items are strongly recommended.

2 A pump rebuild kit is available from Teledyne API's Technical Support including all instructions and required parts

 3 Replace desiccant bags each time the inspection plate for the sensor assembly is removed.

 

  ​

Downloads

Downloads: Sharepoint List not found!

Specifications


Ranges 0 - 5 ppb to 0 - 1 ppm NO
Measurement Units  ppb, ppm, μg/m3, mg/m3 (selectable)
Zero Noise < 20 ppb (RMS)
Span Noise < 0.2% of reading (RMS) above 20 ppt
Lower Detectable Limit   < 40 ppb
Zero Drift < 25 ppb/24 hours
Span Drift < 0.5% of full scale/24 hours
Rise/Fall Time < 30 seconds to 95%
Linearity 1% of full scale
Precision 0.5% of reading above 5 ppb
Sample Flow Rate  900 cc/min ±10%
Power Requirements  80V-100V - 250V-240V,( 50/60 Hz)
Analog Output Ranges 10V, 5V, 1V, 0.1V (selectable)
Recorder Offset ±10%
Included I/O 1 x Ethernet: 10/100Base-T
2 x RS232 (300-115,200 baud)
2 x USB device ports
8 x opto-isolated digital outputs
6 x opto-isolated digital inputs
4 x analog outputs
Optional I/O 1 x USB com port
1 x RS485
8 x analog inputs (0-10V, 12-bit)
4 x digital alarm outputs
Multidrop RS232
3 x 4 - 20mA current outputs
Operating Temperature Range 5 - 40ºC (with US EPA Approvall)
Dimensions (HxWxD) 7” x 17” x 23.5” (178 x 432 x 597 mm)
Weight  33 lbs (15 kg)

Applications

Ambient Air Monitoring

EPA Equivalency

Dilution CEM - Stack and Air Monitoring

 

Description

 

The T500U CAPS NO2 Analyzer represents the next generation of criteria pollutant monitoring technology for the direct measurement of Nitrogen Dioxide (NO2) in air. The instrument utilizes a patented Cavity Attenuated Phase Shift (CAPS) technique to provide an extremely sensitive, fast and accurate NO2 measurement in a cost effective and low maintenance instrument package. 

The T500U provides a direct NO2 measurement, eliminating the need for catalytic conversion or reagents which introduce measurement artifacts inherent in traditional heated metal converter based chemiluminescense instruments. The CAPS measurement approach also reduces maintenance items, and removes the need for hazardous materials and high temperature scrubbers within the instrument. The T500U has an expected maintenance interval of one year under typical ambient monitoring conditions. 

All T Series instruments offer an advanced color display, capacitive touch screen, intuitive user interface, flexible I/0, and built-in data acquisition capability. All instrument set up, control and access to stored data and diagnostic information is available through the front panel, or via RS232, Ethernet, or USB com ports either locally or by remote connection using the included APIcom™ software.

Upon request, the Model T500U is available with NumaView ™ software. NumaView™ Remote PC software allows for a remote connection with virtual interface and data downloading capability to analyzers operating NumaView ™ software.